User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: ME-297 DATE: 16-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 42 NUMBER OF LINES: 129 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5205A CONFIGURATION: Fluke 5220A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F 1.002 DISP For the duration of this calibration, the Test 1.002 DISP Instrument should rest on its back. 1.003 DISP Zero the meter movement using the center screw. 1.004 HEAD {Frequency Response Tests} 1.005 DISP Set Test Instrument fuction to AC VOLTS. 1.006 DISP Using the mini-pin to double banana cable, connect 1.006 DISP the Red pin to the Test Instrument V-[24]-A jack 1.006 DISP and the Black pin to the Test Instrument COM jack. 1.007 5100 2.5 2.5V 4/ 50H 2W 2.001 5100 2.5 2.5V 4/ 1KH 2W 3.001 5100 2.5 2.5V 4/ 10KH 2W 4.001 5100 2.5 2.5V 7/ 20KH 2W 5.001 HEAD {Volts AC Tests} 5.002 5100 2.5 2.5V 4/ 400H 2W 6.001 5100 10 10V 4/ 400H 2W 7.001 5100 50 50V 4/ 400H 2W 8.001 5100 250 250V 4/ 400H 2W 9.001 5100 500 500V 4/ 400H 2W 10.001 ASK- R 10.002 DISP Remove the Test Lead from the Test Instrument V-[24]-A 10.002 DISP jack and install in the 1000V AC/DC 1000[24]/V jack. 10.003 5205 1000 1000V 5/ 400H 2W 11.001 DISP Remove the Test Lead from the Test Instrument 1000V jack 11.001 DISP and install in the 5000V AC/DC 1000[24]/V jack using 11.001 DISP the mini-pin to banana adapter. 11.002 5205 5000 500V 7/ 400H 2W 12.001 HEAD {Volts DC (1000 Ohms/Volt)} 12.002 DISP Set Test Instrument Function to DC VOLTS 1K[24]. 12.003 5205 5000 500V 6/ 2W 13.001 DISP Remove the Test Lead from the Test Instrument 13.001 DISP 5000V AC/DC jack and install in the 13.001 DISP 1000V AC/DC 1000[24]/V jack. 13.002 5205 1000 1000V 3/ 2W 14.001 ASK+ R 14.002 DISP Remove the Test Lead from The Test Instrument 14.002 DISP 1000V jack and install in the V-[24]-A jack. 14.003 5100 500 500V 3/ 2W 15.001 5100 250 250V 3/ 2W 16.001 5100 50 50V 3/ 2W 17.001 5100 10 10V 3/ 2W 18.001 5100 2.5 2.5V 3/ 2W 19.001 HEAD {Volts DC and Linearity Tests (20,000 Ohms/Volt)} 19.002 DISP Set the Test Instrument function to DC VOLTS 20k[24]. 19.003 5100 0.5 0.1V 3/ 2W 20.001 5100 0.5 0.3V 3/ 2W 21.001 5100 0.5 0.5V 3/ 2W 22.001 5100 2.5 2.5V 3/ 2W 23.001 5100 10 10V 3/ 2W 24.001 5100 50 50V 3/ 2W 25.001 5100 250 250V 3/ 2W 26.001 5100 500 500V 3/ 2W 27.001 ASK- R 27.002 DISP Remove the Test Lead from the Test Instrument V-[24]-A 27.002 DISP jack and install in the DC +1000V 20,000 [24]/V jack. 27.003 5205 1000 1000V 4/ 2W 28.001 DISP Remove the Test Lead from the Test Instrument 28.001 DISP 1000V jack and install in the +5000V 20,000[24]/V jack 28.001 DISP using the mini-pin to banana adapter. 28.002 5205 5000 500V 6/ 2W 29.001 HEAD {Current DC Tests} 29.002 DISP Remove the Test Lead from the Test Instrument 5000V jack 29.002 DISP and insert in the 10A jack. 29.003 DISP Set the Test Instrument function to DC MA. 29.004 5220 10 10A 3/ 2W 30.001 ASK+ R 30.002 DISP Remove the Test Lead from the Test Instrument 10A jack 30.002 DISP and install in the V-[24]-A jack. 30.003 5220 2500 2500mA 3/ 2W 31.001 5100 500 500mA 3/ 2W 32.001 5100 100 100mA 3/ 2W 33.001 5100 50 50mA 3/ 2W 34.001 5100 10 10mA 3/ 2W 35.001 5100 2.5 2.5mA 3/ 2W 36.001 5100 0.25 0.25mA 3/ 2W 37.001 HEAD {Resistance Tests} 37.002 ASK- R 37.003 DISP Set Test Instrument Function to OHMS 37.004 DISP Set the Range to OHMS {X1}. 37.005 DISP Short the wires from the Test Instrument together 37.005 DISP at the Test Interface Panel and adjust the OHMS ZERO 37.005 DISP control for a 0 indication on the meter. 37.005 DISP 37.005 DISP Remove the short and set the OHMS INF knob for an 37.005 DISP infinity indication on the meter. 37.006 RESF 10Z +13% -11% 2W 38.001 DISP Set the Range to OHMS {X10}. 38.002 DISP Short the wires from the Test Instrument together 38.002 DISP at the Test Interface Panel and adjust the OHMS ZERO 38.002 DISP control for a 0 indication on the meter. 38.002 DISP 38.002 DISP Remove the short and set the OHMS INF knob for 38.002 DISP an infinity indication on the meter. 38.003 RESF 100Z +13% -11% 2W 39.001 DISP Set the Range to OHMS {X100}. 39.002 DISP Short the wires from the Test Instrument together 39.002 DISP at the Test Interface Panel and adjust the OHMS ZERO 39.002 DISP control for a 0 indication on the meter. 39.002 DISP 39.002 DISP Remove the short and set the OHMS INF knob for 39.002 DISP an infinity indication on the meter. 39.003 RESF 1000Z +13% -11% 2W 40.001 DISP Set the Range to OHMS {X1000}. 40.002 DISP Short the wires from the Test Instrument together at 40.002 DISP the Test Interface Panel. 40.002 DISP Adjust the OHMS ZERO control for 0 reading on the meter. 40.002 DISP Remove the short. 40.003 RESF 10KZ +13% -11% 2W 41.001 DISP Set the Range to OHMS {X10000}. 41.002 DISP Short the wires from the Test Instrument together at 41.002 DISP the Test Interface Panel. 41.002 DISP Adjust the OHMS ZERO control for 0 reading on the meter. 41.002 DISP Remove the short. 41.003 RESF 100KZ +13% -11% 2W 42.001 END